Reference Manual

Table Of Contents
7.3.6 Measured level is occasionally unstable
Figure 7-6: Symptom
A
B
C
D
A. Level
B. Time
C.
Actual level
D. Reported level
Possible cause
The product surface is close to a suppressed false echo.
Recommended actions
If possible, remove the disturbing object.
Related information
Analyze the echo peaks
Adjust the general threshold
Reference Manual Service and troubleshooting
00809-0100-4480 September 2020
Reference Manual 57