7.3.9 Incorrect level when product surface is above 50%
Symptom
The reported level is incorrect when the product surface is above the 50% level.
Figure 7-9: Symptom
A
B
C
D
A. Level
B. Time
C.
Actual level
D. Reported level
Possible cause
A strong double bounce echo is interpreted as the product surface.
Recommended actions
Move the transmitter to another position.
Related information
Mounting position
Service and troubleshooting Reference Manual
September 2020 00809-0100-4480
60 Rosemount 1408H Level Transmitter