Reference Manual

Table Of Contents
Procedure
1. Under
Menu, select Parameter Service Tools Analog Out Calibration.
2. Perform the calibration of 4 mA.
a) Select Enter 4 mA Fixed Current Mode to set the analog out to 4 mA.
b) Measure the analog output with the ampere meter.
c) In the 4 mA Measured Current box, enter the measured current.
d) Select Write to device.
e) Select Calibrate 4 mA.
f) Measure the analog output with the ampere meter and verify it is within
4±0.01 mA.
3. Perform the calibration of 20 mA.
a) Select Enter 20 mA Fixed Current Mode to set the analog out to 20 mA.
b) Measure the analog output with the ampere meter.
c) In the 20 mA Measured Current box, enter the measured current.
d) Select Write to device.
e) Select Calibrate 20 mA.
f) Measure the analog output with the ampere meter and verify it is within
20±0.01 mA.
4. Select Exit Fixed Current Mode.
7.5.6 View input registers
Measured data is continuously stored in the input registers. By viewing the contents of the
input registers, expert users can check that the transmitter works properly.
Procedure
1.
Under Menu, select Diagnosis Read Input Registers.
2. In the Input Reg. Number box, enter the input register to start reading from.
The default start register is 5100 (Alert Summary).
3. Select Write to device.
4. Select Read to device (not required in Rosemount IO-Link Assistant).
Five registers will be read, starting from the selected number.
Reference Manual Service and troubleshooting
00809-0100-4480
September 2020
Reference Manual 67