7.3 HF45 (digital output)
Configuration and full access to all HF45 functions requires a PC with the ROTRONIC HW4 software installed.
Starting with firmware version 1.3, the HF45 offers several communication protocol options that allow users to
read the measurement data without the HW4 software. When using the standard RO-ASCII protocol, access
to some of the HF45 functions is also possible without HW4. For details, see document E-M-AC3000-CP.
IMPORTANT: Depending on the type of digital interface, either the PC or the HF45 must be configured
by the user as indicated below.
a) USB network connection
Prior to connecting the HF45 to a USB port you must install the ROTRONIC USB driver on the HW4 PC. For
instructions see the HW4 manual E-M-HW4v3-Main (§ 7.3)
b) Ethernet (TCP/IP) network connection (wired or wireless)
Prior to connecting the HF45 to an active Ethernet network you must configure the HF45 TCP/IP settings. For
instructions see the HW4 manual E-M-HW4v3-Main (§ 7.4) and technical note E-M-TCPIP-Conf
WARNING: Connecting a device to an active Ethernet network can disrupt communications on the network.
Before connecting the HF4, make sure that it is properly configured for your network
c) RS-485 serial interface (multi-drop)
Instructions for using the HF45 with a RS-485 network are provided in the following manuals:
E-M-HW4v3-Main (§ 7.5), E-M-HW4v3-F2-003 and E-DV04-RS485.01.
Notes:
o Instruments connected to the same RS-485 network must use the same baud rate and each instrument
must be given a unique RS-485 address
o RS-485 Compatibility: The communications protocol used by the HF45 and other AirChip 3000 products
is not compatible with the protocol used by the previous generation of ROTRONIC products. Do not
connect legacy products and AirChip 3000 products to the same RS-485 multi-drop network.
The specifications of the RS-485 interface are as follows:
Baud rate : 19200
Parity : none
Data bits : 8
Stop bits : 1