© 2009-2011; Rotronic AG E-M-HF4-V1_22
Conformity with standards
HF42
HF43
HF45
CE / EMC immunity
EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11
Solder type
Lead free (RoHS directive)
Fire protection class
Corresponds to UL94-HB
FDA / GAMP directives
compatible
Environmental limits
HF42
HF43
HF45
Storage and transit
-50…+70 °C / -20...+70 °C (models with display), 0…100 %RH, non condensing
Operating limits at electronics
-40 … +60 °C / -10….60 °C (models with display), 0…100 %RH, non condensing
Temperature limits at probe
-50…+100 °C (valid for type D)
Maximum humidity at sensor
100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
Maximum air velocity at probe
20 m/s (3,935 ft /min)
Critical environments
Humidity sensor: as per DV04-14.0803.02 - Critical chemicals
10.2 Dew point accuracy
The HF4 can be configured to calculate either the dew point or frost point based on the measurement of
relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and
temperature conditions as shown in the graph below:
Example: at a temperature of 20 ˚C, a frost point value of -25 ˚C is measured with an accuracy of
± 0.5 ˚C or better.