User Manual

E-M-HF4-V1_22
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF4 Humidity Temperature
Transmitters: User Guide
Instruction Manual
Document Type
Page
3 of 29
Document title
© 2009-2011; Rotronic AG E-M-HF4-V1_22
Applicability:
This manual applies to all instruments of the HF4 series with firmware version 1.x, where 1.x can be 1.0, 1.1,
etc. Changes to the last digit of the version number reflect minor firmware changes that do not affect the
manner in which the instrument should be operated.
1 Overview
The HF4 transmitter measures temperature, relative humidity and the dew or frost point. The HF4 series is
designed for fixed installation in HVAC and light industrial applications where a high measurement accuracy is
required at conditions within the range of 0 to 100 %RH and -50 to 100 C (-58 to 212 F). The electronics
operating range is limited to -40…60 °C (-10…60°C with the optional LC display).
The HF4 features well proven sensors and a robust housing. Digital signal processing ensures consistent
product performance and also facilitates the task of field maintenance with features such as potentiometer
free digital calibration.
Based on the ROTRONIC AirChip 3000 digital technology the HF4 series offers the following functions:
User configurable settings
Calculation of the dew or frost point
Humidity temperature calibration and adjustment
Simulator mode
Automatic humidity sensor test and drift compensation
Fail safe mode
Data recording
The ability for the user to easily update the AirChip 3000 firmware means that instruments of the HF4 series
can be kept up-to-date regarding any future functionality improvement.
HF4 transmitters with analog output signals: two types of electronic circuit are available: HF42: 2-wire,
loop powered (4…20 mA current signal) and HF43: 3-wire (voltage or current signal). Both circuit types
provide linear analog outputs signals for transmission over a length of cable to a remote display, recorder,
controller or data processing unit and can be used to measure humidity only, temperature only or both
parameters.
HF4 transmitters with digital output: the HF45 is available with the following digital interface options: RS-
485, USB + RS-485 or Ethernet (TCP/IP) + RS-485. Use of a digital interface allows the simultaneous reading
of relative humidity, temperature and calculated parameter. The communication protocol used by the RS-485
interface offers several options that can be configured by the user. This includes a Modbus compatible
protocol.