Event
Description
Docking station
connected
2
A docking station was connected and communication was established
Logger language file
downloaded
A different internal language file was loaded or the same file was
loaded again
Log function
programmed
The log function has been programmed for input #
Event file deleted
Reserved for the factory
1
connection / removal of analog probes is not recorded
2
only when the docking station has internal electronics
Events triggered from the keypad:
Event
Description
Humidity adjusted
Humidity adjustment of the probe connected to input #
Temperature adjusted
Temperature adjustment of the probe connected to input #
Logging started
Start recording data from input #
Logging stopped
manually
Data recording of input # ended before the programmed stop time
Prior device time
Device date and time prior to change (firmware v1.2a and up)
New device time
Device date and time after change (firmware v1.2a and up)
Log function
programmed
The log function has been programmed for input #
2.11 Operating limits (electronics only)
Models without display:
-10…50°C with the factory supplied alkaline battery
-30…70°C with a lithium battery or an external AC adapter
Models with display:
-10…50°C with the factory supplied alkaline battery
-10…60°C with a lithium battery or an external AC adapter
The HygroClip probes have generally wider operating limits (see individual probe specifications).