The relay contacts are specified as follows:
Contact material
Silver alloy
Initial contact resistance
100 mΩ
Initial insulation resistance
Min. 1,000 MΩ at 500V DC
Nominal switching capacity 1)
2 A 250 VAC
Minimum switching power 1)
1,500 VA
Maximum switching voltage 1)
250 V AC
Maximum switching current 1)
2 A AC
Minimum switching capacity 1)
100 mA, 5 V DC
Expected life (min. operations)
Mechanical: 5 x 10
6
Electrical: N.O. 5 x 10
4
N.C. 3 x 10
4
1) Resistive
The DS-R-1 docking station is provided with two matching 4-pin connectors. The pins of these
connectors are numbered on the connector solder side (see pin-out diagrams).
Relay 1
Not used
Logical Inputs
Relay 2
Not used
RS-485
12VDC Power
(+) center
USB