Manual

IN-E-HL-NT-V2_12
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroLog HL-NT data logger:
instruction manual
Instruction Manual
Document Type
Page
42 of 48
Document title
© 2009-2012; Rotronic AG IN-E-HL-NT-V2_12
EXIT: highlight this menu item and press on ENTER to exit the ADJUST function
The bottom of the display shows the word “Current” and the value read by the probe at the time
the ENTER key was pressed
Note: you can exit the ADJUST function at any time (without adjusting the probe) by pressing the
MENU key.
10.2 Multi-Point adjustment
This is an HW4 software function and is available only when the HygroLog HL-NT is connected to
a PC running the HW4 software (version 2.3.0 or higher).
See the following documents:
o E-M-HW4v3-F2-013: HW4 software version 2: Device Manager HygroLog HL-NT data
logger
o E-M-HW4v3-A2-001: HW4 software version 2: Probe Adjustment function AirChip 3000
devices
11. Warnings and useful tips
Battery Power: when no external AC adapter is connected, use the HW4 software to set the
display to automatically go blank after a specific time interval when there is no keypad activity
or communication over the UART interface.
Set both the display refresh rate and the log interval to the highest values compatible with
your application.
Avoid using a docking station with internal electronics without using an AC adapter.
Do not connect several probes to the HygroLog HL-NT without using an AC adapter.
Do not attempt to charge a regular battery: when using a regular battery, configure the
logger with HW4 so as to disable the internal battery charge function of the logger. Trying to
recharge a regular battery is potentially dangerous.
Date and Time Format: prior to using your logger for the first time, be sure to verify the
following configuration settings: date format (including separator), time format (including
separator) and decimal separator. In order to be able to read log files on a PC, these settings
must be identical to the settings of the PC operating system. If necessary, use HW4 to re-
configure your logger.
Internal Probe (Input 1): for a faster response, use the probe in the extended position.
Compliance with ERES regulatory requirements: Use the HW4 software to disable either
the MENU key or the entire keypad. Disabling the MENU key prevents unauthorized changes
to the configuration of the HygroLog HL-NT and also prevents any probe adjustment.