Specifications

Table 8: Survey of the inspection, maintenance and cleaning work
Inspection Maintenance Cleaning
Rubber gas-
ket of the
vessel lid
Visual inspection be-
for each sterilization
process:
Ensure that the
gasket is intact
before starting!
The rubber gasket is
a wearing part.
It must be renewed
in the event of dam-
age or wear.
Carefully remove
grime with water.
Valves
Visual inspection be-
for each sterilization
process:
Watch for grime!
For safety reasons
the lid must be
returned to the
manufacturer when
the valve has become
clogged.
Do not try to
clean yourself!
Heater
Monthly visual in-
spection:
Watch for scaling!
Eco-friendly descaling
with vinegar at a mod-
erate temperature.
___________
Autoclave
The autoclave must
be checked every
2-3 years by the
manufacturer!
Recalibrate the tem-
perature and pres-
sure indicator yearly.
After sterilization
autoclaves must be
stored in an open
condition.
Monthly cleaning
of the inside de-
pending upon the
frequency of use.
Do not use a
scouring agent!
Sterilization-
process
Check effective-
ness* in conformity
with DIN 58946-8
every 6 month.
* Please contact us in the event of uncertainty. We will be happy to supply you with
any information required.
Check of Temperature and pressure indicator:
The autoclave is to heat up to 1 bar and 2 bar. Than the follow-
ing data have to appear:
Pressure: 1bar Temperature: 121°C
Pressure: 2bar Temperature: 134°C
If there are any deviations the autoclave has to be checked by
the manufacturer or, pressure -, respectively temperature-
indicator has to be replaced.
_____________________________________________________________________
page 30