Specifications

4 Optional extras: Media sensor
The media sensor can only be used in conjunction with the
MCS control (see fig.2 page 23)
WARNING: Do not open the exhaust steam valve after
the sterilization of liquids (media) due to the danger of
delayed boiling. Allow the autoclave to cool down natu-
rally.
Only open the autoclave, when
the chamber temperature has cooled down to at
least 70°C
the autoclave has no remaining pressure (0 bar)
The media sensor is used to establish the actual temperature of
a liquid (medium) that is to be sterilized during the heating-up
and cooling-down stages. The sensor is directly immersed in
the medium (reference container).
1. Activation of the media sensor (see fig. 2 page 23)
a) To activate the media sensor press the "E" key before
the autoclave is started with the "F" key. The LED indi-
cator (6) lights up to confirm that the media sensor has
been switched on.
b) Press the "E" key to call the media temperature during
the sterilization process.
c) NOTE: When a media sensor is being used the sterili-
zation time only commences when the medium has
reached the setpoint temperature.
An additional sterilization time as described under chapter
2.6.2 therefore is not necessary.