Specifications

1 Safety Instructions
„SANOclav" stands for state-of-the-art small-batch autoclaves (small
sterilizers). All safety-technical aspects, in so far as they are within
the manufacturer's sphere of control, have been taken into account.
The series KL and TKL autoclaves are insufficiently equipped
for sterilisation of medical products within the meaning of the
European Guideline 93/42/CCE and are therefore not suitable.
For causes outside the manufacturer's sphere of control, no re-
sponsibility is accepted for consequential damage, injury to the
sterilization product, device, appliance or the surrounding area.
The basic prerequisite for operation ensuring freedom of hazard
to a great degree is that:
the devices are only used by persons who can offer the
guarantee for proper handling on the basis of their training
or their knowledge and practical experience.
These Operating Instructions, as well as any other appro-
priate additional documents, must be made available to the
operating personnel at all times for their work. Operating
personnel are obliged to observe these Operating Instruc-
tions and other documents.
Unqualified personnel are not allowed to work at the auto-
claves or be in their vicinity.
The autoclaves most not be technically changed in any
manner without previously consulting the manufacturer.
These safety instructions do not claim to be complete. In the
event of queries and problems please consult our specialists who
will be happy to help you and supply corresponding information.
__________________________________________________
page 1
2.7.1 Trouble shooting for the MCS control
Table 7: Reason for error message and possible cause
Display Description Cause
Err 1
Operation was aborted during the
heating-up stage
for example: Power failure
Err 2
Operation was aborted during the
sterilization period
for example: Power failure
Err 3
Operation was interrupted Defective (broken) chamber sensor
Err 4
Operation was interrupted Defective (broken) media sensor
Err 5
Operation was interrupted
The chamber or media temperature
dropped by more than 2°C below the
setpoint temperature during a sterili-
zation period of more than 2 minutes
F
- - -
Arises when the device is switched
on or after Err 3 and Err 4
Broken chamber sensor
F
- - -
Arises when the device is switched
on or after Err 3 and Err 4
Short-circuited chamber sensor
Table 8: Action after an error message
Display Action
Err 1
Repeat sterilization. Consult manufacturer if message reappears.
Err 2
Repeat sterilization. Consult manufacturer if message reappears.
Err 3
Autoclave must be repaired by the manufacturer
Err 4
Autoclave must be repaired by the manufacturer
Err 5
Check rubber gasket for proper seat in the lid and then repeat steriliza-
tion. Consult manufacturer if message reappears.
F
- - -
Autoclave must be repaired by the manufacturer
F
- - -
Autoclave must be repaired by the manufacturer
All error messages must be confirmed with the "F" key. This
resets the autoclave to its initial state. The messages "Err 3",
"Err 4" and "F---" cannot be followed by another operation. The
messages "F
---
" or "F
---
" appear when the "F" key (device
ON/OFF) is depressed.
__________________________________________________