Specifications
WARNING: Electromagnetic compatibility:
0
1. To guarantee proper sterilization after a mains
failure, the sterilization must be repeated for the
set time when the power has returned.
2. Do not complete any changes (e.g. a different
power supply cable) without previously consulting
the manufacture as this can adversely influence
the autoclave's immunity to malfunctions.
2 Details concerning use
2.1 Fields of application
The autoclaves of type „KL + TKL..“ are for:
Sterilization in high-pressure steam
Do not sterilize in those autoclaves:
⇒ packed solids
⇒ liquids in closed containers
⇒ further the following dangerous liquids:
• explosive
• high inflammatory
• easy inflammatory
• inflammatory (if the maximum allowed temperature
is above the flamepoint)
• very toxic
• toxic
• effective for burning
__________________________________________________
page 9
C: After sterilization
WARNING: Please observe the „Safety Instructions“
and „Acceptable Risks“ (chapter 1 and 1.2), but par-
ticularly the section „After Sterilization“.
• Turn off the heater when the sterilization time has ended
• The steam pressure control valve can now be carefully
opened and the pressure slowly vented until there is no re-
maining pressure in the chamber (manometer = 0 bar)
• Let the autoclave cool down to at least 80°C
• Before opening the lid check again if there is no pressure
remaining in the chamber
• Now you can open the autoclave and take off the steriliza-
tion products.
2.6.2 Sterilization of liquids
A: Preparation for operation
WARNING: Please observe the "Safety Instructions" and "Ac-
ceptable Risks" (Chapters 1. and 1.2), but particularly the sec-
tion "Preparations for Operation".
The same preparations are required as described in Section
2.6.1 A, except that the following additional and modified pre-
paratory work is necessary:
1. Do place the container holding the liquid to be sterilized deep
but not directly on the autoclave base.
CAUTION: Do not use closed containers. Closed containers
may lead to a retardation of ebullition (delayed boiling).
__________________________________________________
page 18