4.2 Establishing power supply
WARNING!
Using the wrong power packs may cause bodily injury or
damage to the device
Conventional power packs may deliver a higher voltage than is indi-
cated on them. The device may overheat, catch fire, melt or short-
circuit.
Use only the original seca power pack as contained in the scope of
delivery and listed in the section entitled “Spare parts” on page 74.
The connection for the power pack is located on the underside of the weigh-
ing platform. To establish the power supply, proceed as outlined below.
1. Insert the plug required for your power supply into the power pack.
2. Tilt the device forward.
3. Insert the device plug of the power pack into the connector socket of the
device.
4. Carefully return the device to an upright position.
5. Plug the power pack into a power supply socket.
4.3 Setting up the device
The device is fully assembled upon delivery.
NOTICE!
Incorrect measurement due to force shunt
If the device with its housing is in contact with something, e.g. due to
an uneven or soft floor covering, weight will not be measured cor-
rectly.
Set up the device so that only its foot screws are in contact with the
floor.
1. Place the device on a firm, level surface.
NOTICE!
Incorrect alignment may cause incorrect measurements
The spirit level is very sensitive. Additional weights, such as towels,
can result in incorrect scale alignment.
Align the device only without a load on it.
2. Level the device by turning the foot screws.
The air bubble in the spirit level must be located in the center of the circle.