Product Manual

Operation • 31
English
5.3 Performing a bioimpedance measurement
Switching on the device 1. Ensure that the device has no load.
2. Press the ON/OFF button.
The start screen appears.
The
weight/height tab is active.
Determining weight and height Determine weight and height as described in the section entitled
“Determining weight and height” from page 27.
Checking module selection
1. Activate the bia tab.
Bioimpedance analysis is available when the message “Self-test running”
is no longer displayed.
If the device is configured accordingly (see “Configuring the device” on
page 61), the
module selection dialog window appears. The Raw data for
impedance
evaluation module is deactivated at the factory.
Activation/deactivation of the
Raw data for impedance, Energy and Health
risk
evaluation modules affects bioimpedance analysis as outlined in the
table below.
2. Check the module selection.
Evaluation
module
Setting
= activated,
- = deactivated
Effect
a
a.Interrogation of PAL and WC in the Module-specific entries dialog window.
Dialog window is skipped if the Energy
and Health risk evaluation modules
are deactivated.
Energy
Physical activity level (PAL) is interrogated
-
No interrogation of physical activity level
(PAL)
Health risk
Waist circumference (WC) is interrogated
-
No interrogation of waist circumference
(WC)
Raw data for
impedance
- Measurement duration: 20 seconds
Measurement duration: 75 seconds
Raw data for impedance available for
19 frequencies