Product Manual

Function check • 65
English
9. FUNCTION CHECK
Perform a function check prior to each use.
A complete function check includes:
visual inspection for mechanical damage
checking the alignment of the device
visual and function check of the display elements
function check of all the controls shown in the section entitled “Overview”
function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled “What do I do if ...” in this
document.
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled “What do I do if
...” in this document, you may not use the device.
Have the device repaired by seca service or by an authorized service
partner.
Follow the section entitled “Servicing” in this document.
Follow the section entitled “Servicing/recalibration” in this
document.
10.MAINTENANCE/SUBSEQUENT VERIFICATION (VERIFIED MODEL)
10.1 Information on maintenance and subsequent verification
We recommend having your device serviced prior to subsequent verification.
The device’s measurement technology for bioimpedance analysis (BIA) must
be checked every two years.
NOTICE!
Incorrect measurements as a result of poor servicing
Please only have servicing and repairs performed by an authorized
service partner.
You can find a service partner in your area at www.seca.com or by
sending an e-mail to service@seca.com.
Have an authorized technician perform subsequent verification according to
national legal regulations.
Subsequent verification is necessary whenever one or more verification seals
are damaged or the contents of the verification counter no longer match the
number on the applicable verification counter sticker.
10.2 Checking the contents of the verification counter
The scale has been verified. Verifications may only be performed by autho-
rized agencies. To guarantee this, the scale is equipped with a verification
counter which records each change in verification-related data.
If you want to check whether the scale has been properly verified, follow the
instructions below.
1. Switch on the device.
2. Press the
switch menu button.