Product Manual

Troubleshooting • 67
English
12.TROUBLESHOOTING
12.1 Power supply and display
12.2 Height and weight
Problem Cause Remedy
Device cannot be switched on
No power supply Check whether power is being supplied
Power pack faulty Replace power pack with original part
Touchscreen display stays
dark
Device on standby
Touch the touchscreen display
Press the on/off button
Place a load on the device
Device not switched on Switch on device
No power supply Check whether power is being supplied
Touchscreen display faulty Inform seca service
Touchscreen display not
reacting
Device in undefined state following
implausible input
Switch off the device (hold the on/off
button for approx. 3 seconds)
Switch on the device again
Image on touchscreen display
faulty
Touchscreen display faulty Inform seca service
Problem Cause Remedy
0.00 does not appear before
weighing
Device had load on it before it was
switched on
Remove the load from the device
Switch off the device, then switch back on
again
The STOP message appears Maximum load exceeded Remove the load from the device
TEMP message appears
Ambient temperature too high or too
low
Set up the device in an ambient tempera-
ture between +10 °C and +40 °C
Wait for around 15 minutes until the
device has adapted to ambient
temperature
The message ER11 appears
Device overloaded as a whole or in one
corner
Remove the load or distribute the load
evenly
Restart the device
The message ER12 appears
Device switched on with too heavy a
load
Remove the load from the device
Restart the device
The message ER16 appears
Natural vibration has been induced in
device, zero point could not be
determined
Restart the device
Restart measurement