12.3 Bioimpedance analysis
Problem Cause Remedy
bia tab activated, but module
selection does not appear
Module selection deactivated
Check setting and change if necessary
(see “Adapting default module selection for
bioimpedance analysis” on page 61)
Not all modules activated in
module selection
Default module selection specified in
which some modules are deactivated
Activate missing modules directly in the
module selection and perform the
measurement
Adapt default module selection (see
“Creating default module selection” on
page 62)
The following message
appears: “Electrode detection
failed.”
Patient’s skin too dry
Spray the skin with electrode spray at the
contact points
Patient’s skin too calloused
Spray the skin with electrode spray at the
contact points
Electrodes faulty Inform seca service
No PAL value can be entered
after bioimpedance
measurement
Energy
evaluation module deactivated
If the
Energy evaluation module is not
required, continue and complete
measurement
If the
Energy evaluation module is
required, activate the evaluation module
(see “Checking module selection” on
page 31 or “Creating default module
selection” on page 62)
No waist circumference can
be entered following
bioimpedance measurement
Health risk
evaluation module
deactivated
If the
Health risk evaluation module is not
required, continue and complete
measurement
If the
Health risk evaluation module is
required, activate the evaluation module
(see “Checking module selection” on
page 31 or “Creating default module
selection” on page 62)
Results of bioimpedance
measurement deviate
significantly from expected
results
Patient moved during the
measurement
Request that the patient does not move
during the measurement and repeat the
measurement
Patient used different pairs of hand
electrodes on the left and right
Ensure that the patient uses the same
hand electrodes on both sides and repeat
the measurement
Electrodes faulty Inform seca service
Value of an evaluation
parameter is shown in red
Value outside the normal range
determined for the evaluation
parameter
Repeat the measurement in order to
exclude measurement errors
If the repeat measurement also produces
a value outside the normal range, take the
value into account in further investigations
After a different tab has been
called up temporarily, the
assigned seca patient file is
no longer shown in the patient
tab
seca patient file selected, but selection
not confirmed
Re-assign seca patient file and press the
confirm button (see “Assigning a
measurement to a patient file” on page 36),
only then call up another tab