Product Manual

Technical data • 73
English
13.3 Weighing data (verified model)
13.4 Weighing data (unverified model)
seca 515
Accuracy class in accordance with Directive 2014/31/EU III
Measuring method 4 load cells
Maximum load
Partial weighing range 1
Partial weighing range 2
150 kg
300 kg
Minimum load
Partial weighing range 1
Partial weighing range 2
1 kg
2 kg
Increments
Partial weighing range 1
Partial weighing range 2
50 g
100 g
Tare range to 300 kg
Accuracy on initial verification
Weighing range 1: 0 to 25 kg
Weighing range 1: 25 kg to 100 kg
Weighing range 1: 100 kg to 150 kg
Weighing range 2: 0 to 50 kg
Weighing range 2: 50 kg to 200 kg
Weighing range 2: 200 kg to 300 kg
±25 g
±50 g
±75 g
±50 g
±100 g
±150 g
seca514
Measuring method 4 load cells
Maximum load
Partial weighing range 1
Partial weighing range 2
150 kg / 330 lbs / 24 sts
300 kg / 660 lbs / 47 sts
Minimum load
Partial weighing range 1
Partial weighing range 2
1.0 kg
2.0 kg
Increments
Partial weighing range 1
Partial weighing range 2
50 g / 0.1 lbs
100 g / 0.2 lbs
Tare range 300 kg
Accuracy
•0 to 35 kg
35 kg to maximum load
•0 to 75 lbs
75 lbs to maximum load
0 to 5.5 sts
5.5 sts to maximum load
±100 g
±0.3%
±0.2 lbs
±0.3%
±0.2 lbs
±0.3%