LS3200 POINT LEVEL SWITCH CALIBRATION
REFERENCE DRAWING
5.3 Calibration - Level
Using Figure 10.0 as a location guide adjust the system as follows:
1. Remove the instrument enclosure lid by turning ccw.
2. Apply power to the unit. Allow 5 minute warm-up.
3. For optimum calibration results, wet sensor and drain but do not dry.
4. Ensure that the tank liquid level is below the probe sensor tips.