- SMC Neurovascular Array Coil Operator's Manual

12
Rev 05
Connecting
the Cable
Attach the Synergy MultiConnect (SMC) holder to the same side of the support as the
patient ineterface control unit (PICU) facing the magnet. In this example, the PICU is on
the left side. Slide the holder toward the bore, even with the patients knees.
For systems with PICU on left side of magnet, transfer SMC connector across table
to the PICU. IMPORTANT: For SMC cable to reach PICU on left side of magnet,
patient will need to be in position within the magnet, e.g., coil at magnet isocenter.
Route the coil cable and connector to the SMC holder, as shown above.
Join the SMC and HNC connectors, and place the assembly into the SMC holder, as
shown above.
Landmark on the anatomy to be imaged and perform TRAVEL-TO-SCANPLANE.