12. INSPECTION & ADJUSTMENT
The instrument has passed the procedure of inspection and adjustment before releasing to the market,
which ensures that it meets quality requirement. However, after long periods of transportation or the
changeable environment, some influences may occur to the internal structure.
Therefore, before the instrument is used for the first time, user should check and adjust the functions
we introduced in this session to ensure the precision of the job.
12.1 PLATE VIAL
Inspection
Loosen the horizontal tangent screw, rotate the
equipment to ensure that the plate vial is parallel
to the direction of foot screw AB. Adjust the
screw A/B in opposite direction to move the
bubble in the center. Rotate the instrument to
180° to see whether the bubble is in center, if not,
the plate bubble needs to be adjusted.
Adjustment
1. If the bubble of the plate vial moves away
from the center, bring it half way back to the
center by adjusting the screws, which is parallel
to the plate vial. Correct the remaining half by
adjusting pin.
2. Rotate the instrument in 180º to check whether
the bubble is in the center. If not, repeat Step 1.
3. Rotate the instrument in 90º, adjust the third
screw. Repeat the steps until the bubble remains
in the center in any direction.