12.2 CIRCULAR VIAL
Inspection
It is not necessary to adjust the circular vial,
except the bubble is not in the center after the
adjustment of plate vial.
Adjustment
If the bubble of the circular vial is not in the
center, adjust the bubble to the center by using
the adjusting pin or hexagon wrench.
First, loosen the screw opposite to the offset side,
and then tighten the other adjusting screw on
the offset side, bringing the bubble to the center.
When the bubble stays in the center, keep the
tightness of the three screws uniformly.
12.3 TILT-SENSOR
Adjust the plate vial at first, leveling the equipment. Then enter the adjustment page of electronic
bubble. Press [setting] to set the sensor as 0 in the current state.