8.4 Lighng maintenance:
DANGER!
Maintaining lights and indicators is a safety crical task. If your
lights and indicator system shows a defect, please contact your
Sunrise Medical Approved Dealer.
All lights and indicators are state of the art LED low energy, no
maintenance units. There are no light bulbs used. The high
intrinsic reliability of these units means that they are extremely
unlikely to fail under normal usage. If a failure should occur,
(may be as a result of an impact), the damaged light assembly
must be replaced in its enrety. The individual LED’s cannot be
replaced.
CAUTION!
• We recommend the use of Sunrise Medical authorized spare
parts only.
• Be advised that all lighng circuits are electronically
protected. In the event of a short circuit current will be
limited to a safe level. Once the fault is removed the system
resets automacally.
8.5 Cleaning and disinfecon
The wheelchair should be wiped over once per week with a
slightly damp, not wet, cloth and any u or dust that has
accumulated around the motors should be blown or dusted
away.
CAUTION!
Make sure that you dry all parts of your wheelchair if it
becomes wet or damp aer cleaning or if it is used in a wet or
damp atmosphere.
DANGER!
It is important that should the wheelchair be used by more than
one person it is cleaned thoroughly to ensure there is no cross
infecon.
Hygiene measures when being re-used:
Prior to the wheelchair being re-used, it must be carefully
prepared. All surfaces which come into contact with the user
must be treated with a disinfecon spray.
To do this, you must use a suitable disinfectant for rapid alcohol-
based disinfecon of medical products and medical devices.
Please take into account the manufacturer’s instrucons for the
disinfectant you are using.
CAUTION!
• Do not use solvents, bleaches, abrasives, synthec
detergents, wax polishes or aerosols.
• Disinfectants may be used in diluon as specied by their
manufacturer.
• Ensure surfaces are rinsed with clean water and dried
thoroughly.
WARNING!
• Always read the label on any commercial or domesc
cleaning substances.
• Always follow the instrucons carefully.
Cleaning controls: & Controller
Should the control of your wheelchair become soiled or dirty,
it can be wiped with a damp cloth with a dilute disinfectant
unl clean.
DANGER!
Important: If the wheelchair is to be used by more than one
person, follow the cleaning and disinfecon instrucons
carefully to ensure that there is no cross infecon.
8.6 Medium to long term storage:
When storing your wheelchair for long periods of me (in
excess of one week), follow these simple instrucons:
• Fully charge the wheelchair for at least 24 hours.
• Disconnect the charger.
• Disconnect the baeries.
WARNING!
Never store your wheelchair:
• Outside.
• In direct sunlight, (plasc parts may discolour).
• Near a source of direct heat.
• In a damp environment.
• In a cold environment.
• With the baeries/baery boxes connected, (even if the
controller is switched o).
Avoiding all of the above will minimise baery deep cycle
discharge and extend baery lifeme.
When returning the wheelchair to use, please reconnect the
baeries/baery boxes and charge the wheelchair for at least
24 hours before use.