User Guide

Teledyne API M100E Analyzer Operation Manual Operating Instructions
101
6.9.4.1. ANALOG OUTPUT SIGNAL TYPE AND RANGE SPAN SELECTION
To select an output signal type (DC Voltage or current) and level for one output channel, activate the ANALOG
I/O CONFIGURATION MENU from the DIAG Menu (refer to Figure 6-17), then press:
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO OUTPUT RANGE: 5V
0.1V 1V 5V 10V CURR ENTR EXIT
These keys set
the signal level
and type of the
selected channel
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
DIAG AIO OUTPUT RANGE: 10V
0.1V 1V 5V 10V CURR ENTR EXIT
Figure 6-20: DIAG – Analog I/O Configuration Menu