User Guide

Operating Instructions Teledyne API M100E Analyzer Operation Manual
102
6.9.4.2. ANALOG OUTPUT CALIBRATION MODE
The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is configured
for automatic calibration of all channels. Manual calibration should be used for the 0.1V range or in cases where
the outputs must be closely matched to the characteristics of the recording device. Outputs configured for
automatic calibration can be calibrated as a group or individually. Calibration of the analog outputs needs to be
carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or
whenever re-calibration is required.
To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU from the DIAG Menu
(refer to Figure 6-17), then press:
DIAG AIO AUTO CALIBRATING CONC_OUT_1
AUTO CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
Exit at any time
to return to the
main DIAG
menu
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO
A
OUTS CALIBRATED: NO
< SET SET> CAL EXIT
If AutoCal has been
turned off for any
channel, the message
for that channel will be
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
If any of the channels have
not been calibrated this
message will read NO.
DIAG AIO AOUTS CALIBRATED: YES
< SET SET> CAL EXIT
Figure 6-21: DIAG – Analog Output Calibration Mode
04515F DCN6048