User Guide

Teledyne API M100E Analyzer Operation Manual Operating Instructions
107
6.9.4.4. ANALOG OUTPUT OFFSET ADJUSTMENT
Some analog signal recorders require that the zero signal to be significantly different from the baseline of the
recorder in order to record slightly negative readings from noise around the zero point. This can be achieved in
the M100E by defining a zero offset, a small voltage (e.g., 10% of span), which can be added to the signal of
individual output channels by activating the ANALOG I/O CONFIGURATION MENU from the DIAG Menu (refer
to Figure 6-17), then press:
DIAG AIO RECORD OFFSET: 0 MV
+ 0 0 0 0 ENTR EXIT
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EXIT
Set the recorder
offset (in mV) of
the selected
channel
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
DIAG AIO CONC_OUT_2 REC OFS: 0 mV
< SET SET> EDIT EXIT
Pressing ENTR accepts the
new setting and returns to the
previous menu.
Pressing EXIT ignores the new
setting and returns to the
previous menu.
Press SET> to select the
analog output channel to
be configured. Then press
EDIT to continue
DIAG AIO CONC_OUT_2 OVERRANGE: ON
SET> EDIT EXIT
Figure 6-26: Analog Output – Offset Adjustment
04515F DCN6048