User Guide

Remote Operation and Advanced Features Teledyne API M100E Analyzer Operation Manual
134
1. With NO power to the instrument, remove its top cover and locate JP2 on the multidrop PCA, which is
assembled with a shunt that jumpers Pins 21 22 (Figure 7-12).
2. Remove a
nd store the shunt (place the shunt on one pin only) for all instruments in the network except the
instrument that is to be the last: make sure a shunt is in place connecting Pins 21 22 for the last
instrument.
Note: If you are adding an instrument to the end of a previously configured chain, remove the shunt
between Pins 21 22 of JP2 on the multidrop PCA in the instrument that was previously the last
instrument in the chain.
3. Close the instrument.
4. Using straight-through, DB9 male DB9 Female cable, interconnect the host and the analyzers as shown
in Figure 7-13.
5. BEFORE co
mmunicating from the host, power on the instruments and check that the Machine ID code is
unique for each. (On the front panel menu, use SETUP>MORE>COMM>ID. Note that the default ID is
typically the model number; to change the 4-digit identification number, press the key below the
corresponding digit to be changed).
NOTE:
Teledyne API recommends setting up the first link, between the Host and the first analyzer and testing it
before setting up the rest of the chain.
Analyzer Analyzer Analyzer Last Analyzer
Female DB9
Male DB9
RS-232
COM2
RS-232
COM2
RS-232
COM2
RS-232
COM2
Host
RS-232 port
Make Sure
Jumper between
JP2 pins 21
22
is installed.
KEY:
Figure 7-13: RS232-Multidrop PCA Host
/Analyzer Interconnect Diagram