User Guide

Teledyne API M100E Analyzer Operation Manual Calibration Procedures
179
8.3. MANUAL CALIBRATION CHECKS
Informal calibration checks will only evaluate the analyzer’s response curve, but do not alter it. It is
recommended as a regular maintenance item, to perform calibration checks in order to monitor the analyzer’s
performance. To carry out a calibration check rather than a full calibration, perform the following procedures:
STEP ONE: Connect the sources of zero air and span gas as shown in Figure 8-1.
STEP TWO: Perform
the zero/span calibration check procedure:
SAMPLE STABIL=XXX.X PPB SO2=XXX.X
< TST TST > CAL SETUP
ACTION:
Record the SO
2
concentration
reading.
A
CTION:
Supply span gas to the instrument
SAMPLE STABIL=XXX.X PPB SO2=XXX.X
< TST TST > CAL SETUP
ACTION:
Record the SO
2
concentration
reading.
Wait until
STABIL is
below 0.5 ppb.
This may take
several minutes.
The value of
STABIL may jump
significantly.
Wait until it falls
below 0.5 ppb. This
may take several
minutes.
SAMPLE STABIL=XXX.X PPB SO2=XXX.X
< TST TST > CAL SETUP
The SPAN key appears during the transition from zero to
span. You may see both keys.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
Scroll the display to the
STABIL test function.
A
CTION:
Supply the instrument with zero gas.
Figure 8-4: Setup for Manual Calibration Checks
04515F DCN6048