User Guide

Specifications, Approvals and Warranty Teledyne API M100E Analyzer Operation Manual
20
2.2. EPA EQUIVALENCY DESIGNATION
The M100E Analyzer is designated as Reference Method Number EQSA-0495-100 as per 40 CFR Part 53 when
operated under the following conditions:
Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
Ambient temperature range of 5
o
C to 40
o
C.
Line voltage range of 100-120 VAC or 220-240 VAC, at 50 or 60 Hz.
Sample filter: Equipped with PTFE filter element in the internal filter assembly.
Sample flow of 650 +/- 65 cm
3
/min.
Vacuum pump (internal) capable of 14"Hg Absolute pressure @ 1 slpm or better.
Software settings:
Dynamic span OFF
Dynamic zero OFF
Dilution factor OFF
AutoCal ON or OFF
Dual range ON or OFF
Auto-range ON or OFF
Temp/Pressure compensation ON
Under the designation, the analyzer may be operated with or without the following optional equipment:
Rack mount with chassis slides.
Rack mount without slides, ears only.
Zero/span valve options.
Internal zero/span (IZS) option with either:
SO
2
permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.
SO
2
permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. Under the designation, the
IZS option cannot be used as the source of calibration.
4-20mA isolated analog outputs.
Status outputs.
Control inputs.
RS-232 output.
Ethernet output.
Zero air scrubber.
4-20mA, isolated output.
04515F DCN6048