User Guide

Theory Of Operation Teledyne API M100E Analyzer Operation Manual
254
A)-HG(" 29.92
_SLOPESAMP_PRESSA)-HG(" SSURESAMPLE_PRE
PRESSCO
(Equation 11-6)
SAMPLE-PRESSURE: The ambient pressure of the sample gas as measured by the instrument’s sample
pressure sensor in “Hg-A.
SAMP_PRESS_SLOPE: Sample pressure slope correction factor. The default setting for Section 6.8 describes
the method for enabling/disabling the TPC feature.
11.7.4. INTERNAL DATA ACQUISITION SYSTEM (IDAS)
The iDAS is designed to implement predictive diagnostics that stores trending data for users to anticipate when
an instrument will require service. Large amounts of data can be stored in non-volatile memory and retrieved in
plain text format for further processing with common data analysis programs. The iDAS has a consistent user
interface in all Teledyne API instruments. New data parameters and triggering events can be added to the
instrument as needed.
Depending on the sampling frequency and the number of data parameters the iDAS can store several months of
data, which are retained even when the instrument is powered off or a new firmware is installed. The iDAS
permits users to access the data through the instrument’s front panel or the remote interface. The latter can
automatically download stored data for further processing. For information on using the iDAS, refer to Section
7.2.
04515F DCN6048