APPENDIX A-2: Setup Variables For Serial I/O, Software Version G.4
Table A-1: Setup Variables, Software Version G.4
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
Low Access Level Setup Variables (818 password)
DAS_HOLD_OFF Minutes 15 0.5–20 Duration of DAS hold off period.
TPC_ENABLE — ON OFF, ON
ON enables temperature and pressure
compensation; OFF disables it.
50 RCELL_SET ºC
Warnings:
45–55
30–70
Reaction cell temperature set point and
warning limits.
50 IZS_SET
1
ºC
Warnings:
45–55
30–70
IZS temperature set point and warning
limits.
DYN_ZERO — OFF OFF, ON
ON enables contact closure dynamic
zero; OFF disables it.
DYN_SPAN — OFF OFF, ON
ON enables contact closure dynamic
span; OFF disables it.
CONC_PRECISION — 3 AUTO,
0,
1,
2,
3,
4
Number of digits to display to the right
of the decimal point for concentrations
on the display. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
REM_CAL_DURATION
17
Minutes 20 1–120
Duration of automatic calibration
initiated from TAI protocol.
CLOCK_ADJ Sec./Day 0 -60–60 Time-of-day clock speed adjustment.
Medium Access Level Setup Variables (929 password)
LANGUAGE_SELECT — ENGL ENGL,
SECD,
EXTN
Selects the language to use for the
user interface. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
MAINT_TIMEOUT Hours 2 0.1–100
Time until automatically switching out
of software-controlled maintenance
mode.
CONV_TIME 33 MS 33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT and UV
detector channels. Enclose value in
double quotes (“) when setting from the
RS-232 interface.
DWELL_TIME Seconds 1 0.1–10 Dwell time before taking each sample.
FILT_SIZE Samples 240,
30
3
,
480
12
1–1000 Moving average filter size.
04515F DCN6048