Q: How do I perform a leak check?
A: Refer to Section 10.3.6.
Q: How do I measure the sampl
e flow?
A: Sample flow is measured by attaching a calibrated flow meter to the sample inlet port when the instrument is
operating. The sample flow should be 650 cm³/min 10%. Section 10.3.6 includes detailed instructions on
perfo
rming a
check of the sample gas flow.
Q: How often do I need to change the particulate filter?
A: Once per week. Table 10-1 contains a maintenance schedule listing the most important, regula
r maintenance
tasks.
Q: What is the averaging time for an M100E?
A: The default averaging time, optimized for ambient pollution monitoring, is 240 seconds for stable
concentrations and 20 seconds for rapidly changing concentrations; Refer to 11.7.1 for more information.
Q: My analyzer has the opt
ional, user -configurable analog output channels. How do I program and use them?
A: Instructions for this can be found in the Manual Addendum for Configurable Analog Output, PN 06270.
Q: How long does the sample pump last?
A: The sample pump should last about one year and the pump diaphragms should to be replaced annually or
when necessary. Use the PRES test function displayed via the front panel to see if the diaphragm needs
replacement (refer to Section 12.1.2).
Q: Do I need a strip
chart recorder or external data logger?
A: No, the M100E is equipped with a very powerful internal data acquisition system. Section 7.2 describes the
setup and op
eration in detail.