Teledyne API M100E Analyzer Operation Manual Optional Hardware and Software
55
Figure 5-2: Current Loop Option Installed on the Motherboard
5.4.1. CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD
VOLTAGE OUTPUTS
NOTE
Servicing or handling of circuit components requires electrostatic discharge protection, i.e. ESD
grounding straps, mats and containers. Failure to use ESD protection when working with electronic
assemblies will void the instrument warranty.
Refer to Chapter 13 for more information on preventing ESD damage.
To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation:
1. Turn off power to the analyzer.
2. If a recording device was connected to the output being modified, disconnect it.
3. Remove the top cover
Remove the set screw located in the top, center of the rear panel
Remove the screws fastening the top cover to the unit (four per side).
Lift the cover straight up.