User Guide

Operating Instructions Teledyne API M100E Analyzer Operation Manual
78
6.3. CALIBRATION MODE
6.3.1. CALIBRATION FUNCTIONS
Pressing the CAL key switches the M100E into calibration mode. In this mode, the user can calibrate the
instrument with the use of calibrated zero or span gases.
If the instrument includes either the zero/span valve option or IZS option, the display will also include CALZ and
CALS keys. Pressing either of these keys also puts the instrument into multipoint calibration mode.
The CALZ key is used to initiate a calibration of the zero point.
The CALS key is used to calibrate the span point of the analyzer. It is recommended that this span
calibration is performed at 80% of full scale of the analyzer’s currently selected reporting range.
Because of their critical importance and complexity, calibration operations are described in detail in other
sections of the manual:
Chapter 8 details basic calibration and calibration check operations.
Cha
pter 9 des
cribes how to perform an EPA protocol calibration.
For more info
rmation concerning the zero/span, zero/span/shutoff and IZS valve options, refer to Section 5.7.
04515F DCN6048