User Guide

Operating Instructions Teledyne API M100E Analyzer Operation Manual
84
In order to compensate for CPU clocks which run fast or slow, there is a variable to speed up or slow down the
clock by a fixed amount every day. To change this variable, press:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS: 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
3x EXIT returns
to the main SAMPLE display
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 0 ) DAS_HOLD_OFF=15.0 Minutes
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X 7) CLOCK_ADJ=0 Sec/Day
PREV JUMP EDIT PRNT EXIT
SETUP X.X CLOCK_ADJ:0 Sec/Day
+ 0 0 ENTR EXIT
SETUP X.X 8) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Continue to press NEXT until …
SAMPLE ENTER SETUP PASS: 818
8 1 8 ENTR EXIT
Figure 6-8: SETUP – Clock Speed Variable
04515F DCN6048