User Guide

Operating Instructions Teledyne API M100E Analyzer Operation Manual
96
6.9.1. ACCESSING THE DIAGNOSTIC FEATURES
To access the DIAG functions press the following keys:
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
From this point
forward, EXIT returns
to the
SECONDARY
SETUP MENU
EXIT returns
to the PRIMARY
SETUP MENU
DIAG PRESSURE CALIBRATION
PREV NEXT ENTR EXIT
DIAG OPTIC TEST
PREV NEXT ENTR EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG ANALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG TEST CHAN OUTPUT
PREV ENTR EXIT
DIAG ELECTRICAL TEST
PREV NEXT ENTR EXIT
DIAG LAMP CALIBRATION
PREV NEXT ENTR EXIT
EXIT returns
to the main
SAMPLE
display
DIAG FLOW CALIBRATION
PREV NEXT ENTR EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
Figure 6-17: DIAG Menu
04515F DCN6048