Now use the 4 BIT-TO-HEXADECIMAL table above to convert this 4 bit value combination into a
two place hexadecimal value:
<fmt> Two Place Hexadecimal Value : 8b
From the OPTIONS <opt> table above, you identify your desired columns with its corresponding
bit value:
<opt> Descriptions <opt> Bit Value
Calculations Errors 0000 0100
Heater State 0000 0010
<opt> 4 Bit Value Combination: 0000 0110
Again, use the 4 BIT-TO-HEXADECIMAL table to convert this 4 bit value combination into a two
place hexadecimal value:
<opt> Two Place Hexadecimal Value : 06
Conclusion: To have the Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak
Hydrogen Reading, Calculations Errors and Heater State columns continuously reported, you
will implement the G serial command (Go Command, refer to LEVEL 1 COMMANDS table) as
follows at the H2scan:
command prompt: G < fmt> <opt> = g 8b 06
Display User Response
H2scan: Typeg 8b 06