Data Sheet

THM 6 Series, 6 Watt
http://www.tracopower.com Page 3 of 4
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
General Specifications
Capacitive load –Single output 3.3 VDC models: 2‘100 F max.
5.0 VDC models: 1‘500 F max.
12 VDC models: 260 F max.
15 VDC models: 210 F max.
24 VDC models: 75 F max.
–Dual output ±5 VDC models: 860 F max. (each output)
±12 VDC models: 150 F max. (each output)
±15 VDC models: 110 F max. (each output)
Temperature ranges – Operating (designed for) –40°C to +88°C (without derating)
– Rated according to IEC/EN 60601-1 –40°C to +70°C (without derating)
– Case temperature +105°C max.
– Storage temperature –55°C to +125°C
Thermal impedance 18 K/W
Humidity (non condensing) 5 % to 95 % rel H max.
Isolation voltage – to meet ES/IEC/EN 60601-1 5000 VACrms, rated for 250 VACrms working
(50Hz, 60sec) voltage, 2  MOPP
Clearance/creepage 8 mm min.
Leakage current ( at 240 VAC, 60 Hz) 2 A max.
Isolation capacitance (input/output) 17 pF max.
Altitude during operation 5000 m
Temperature coefficient ±0.02 %/K typ.
Reliability, calculated MTBF (MIL-HDBK-217F at +25°C, ground benign) 4‘700‘000 h
Switching frequency 250 kHz ±25 kHz (pulse width modulation)
Vibration and thermal shock resistance according to MIL-STD-810F
Safety standards/approvals – Medical equipment ANSI/AAMI ES60601-1:2005/(R)2012,
IEC/EN60601-1 3rd edition
– Certification documents www.tracopower.com/products/overview/thm6
Environmental compliance Reach www.tracopower.com/products/reach-declaration.pdf
RoHS RoHS directive 2011/65/EU
Physical Specifications
Casing material non-conductive black plastic
Base material non-conductive black plastic
Potting material silicone (UL94 V-0 rated)
Package weight 14 g (0.48oz)
Soldering temperature max. 265°C / 10 sec
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.