User`s guide

User's Guide _______________________________________________________________________
36 ___________________________________________________________________ M211371EN-D
Show Firmware Version
VERS<cr>
Example:
vers
DPT145 1.0.23
Show Serial Number
SNUM<cr>
Example:
snum
Serial number : F4040007
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.
R<cr>
Example (measurement message in default format):
r
Tdf= 12.5 'C Tdfatm= 13.6 'C H2O= 15561 ppm P= 0.947 bara
Pnorm= 0.943 bara Rhoo= 1.1 kg/m3 T= 24.5 'C
Tdf= 12.5 'C Tdfatm= 13.6 'C H2O= 15563 ppm P= 0.947 bara
Pnorm= 0.942 bara Rhoo= 1.1 kg/m3 T= 24.5 'C
Tdf= 12.5 'C Tdfatm= 13.5 'C H2O= 15555 ppm P= 0.948 bara
Pnorm= 0.943 bara Rhoo= 1.1 kg/m3 T= 24.5 'C
...
Outputting the results continues in intervals issued with the command
INTV. You can stop the output by pressing the Esc key, or with the
command S. Since the interface is half-duplex, you must enter the
commands when the transmitter is not outputting.
Stop Measurement Output
You can stop the measurement output by pressing ESC, or with the S
command:
S<cr>