User`s guide

Chapter 7 ____________________________________________________________ Troubleshooting
VAISALA ________________________________________________________________________ 51
Error Messages
The transmitter has error codes that can be displayed on the serial line
using the ERRS command. See section Show Currently Active Errors on
page 35.
Table 11 Error Messages
Error Message
Possible Cause and Solution
T MEAS error
MPS1 sensor damaged or missing.
Contact a Vaisala Service Center.
F meas error
Voltage error
Supply voltage out of range.
Check and correct.
Voltage too low error
Program flash check sum error
Internal transmitter error. If the error
remains after a reset and restoring the
factory settings, contact a Vaisala
Service Center.
Parameter flash check sum error
INFOA check sum error
SCOEFS check sum error
CURRENT check sum error