Terminal Application Settings
You need a terminal application to be able to use the service port commands of
the HMDW110 series transmitter. You can download the PuTTY terminal
application from www.vaisala.com or use a terminal application of your choice.
Figure 12 PuTTYTerminal Application
Before starting a terminal session, you must set the following parameters:
n Serial line settings. The default serial line settings of a HMDW110 series
transmitter are 19200 N 8 1:
o
19200 bits per second
o
No parity
o
Eight data bits
o
One stop bit
n Number of the virtual COMport that has been created for your cable by the
Vaisala USBdriver. You can check which port the USB cable is using with the
Vaisala USB Device Finder application that has been installed in the
Windows Start menu. You can also access the USBfinder directly from the
Vaisala-supplied PuTTYapplication by clicking on the USBFinder... button
on the Serial &USBtab.
The default serial line settings are needed for compatibility with
the RDP100 remote display panel. If you are not using the
display, you can change the settings using the seri command.
Note that the service port settings will also change.