Owner manual

Command Description Page
crhclr
Clear user calibration for humidity measurement.
Factorycalibration remains.
37
ct
User calibration for T measurement. 37
ctclr
Clear user calibration for temperature measurement. Factory
calibration remains.
37
fcrh
Two-point calibration after humidity sensor change. 38
l
Show adjustment offset and gain. 38
li
Set adjustment offset and gain. 39
Other commands
filt
Show or set measurement filtering. 39
frestore
Restore factory settings. Clears all user settings, factory
calibration remains.
40
reset
Reset the device. 40
Device Information and Status
Syntax Description
?<cr>
Show listing of device information.
??<cr>
Show listing of device information even if
device is in poll mode and connection has not
been opened using the
open
command.
Example:
?
HMD112 / 1.0.6
Serial number : H0134007
Batch number : H0130002
Sensor number : H0090003
Sensor model : Humicap 180R
Order code : HMD1102A1VA1
Cal. date : 20140619
Cal. info : VAISALA/HEL
Time : 01:43:43
Serial mode : STOP
Baud P D S : 19200 N 8 1
Output interval: 5 S
Serial delay : 25
Address : 0
Filter : 1.000
Ch1 output : 4 ... 20 mA
Ch2 output : 4 ... 20 mA
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
Table 5 ? Command
25
4 Service Port