Owner manual

Syntax Description
amode<cr>
Show currently set analog output mode.
Example:
amode
Ch1 output : 4 ... 20 mA
Ch2 output : 4 ... 20 mA
Table 25 Amode Command
Syntax Description
aover<cr>
Show current aover status.
aover [on|off]<cr>
Set analog output overrange to:
on = Allow output to exceed the scaled range
by 10%.
off = Keep the analog output always in the
scaled range. Note that error state behavior
overrides this setting.
Example (enable 10% overrange for analog outputs):
aover on
AOVER : ON
Table 26 Aover Command
34
4 Service Port