Owner manual

Syntax Description
fcrh<cr>
Start the two-point humidity calibration and
adjustment sequence. If you have changed
the humidity sensor of the device yourself,
you must perform a two-point humidity
calibration and adjustment using this
command. Follow the procedure in section
Two-Point Humidity Calibration and
Adjustment using a computer and a HMK15
Humidity Calibrator on page46
.
Example:
fcrh
RH : 11.3143 1. ref ? 11.3
Press any key when ready ...
RH : 75.0012 2. ref ? 75.4
OK
Table 33 Fcrh Command
Syntax Description
l<cr>
Show the current offset and gain parameters
for user adjustment. This command is useful
for checking if user adjustments have been
applied for humidity and temperature
measurement.
Example (shows default state without user adjustments - offset is 0 and gain is 1 for both
humidity (Cp offset and gain) and temperature):
l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Table 34 L Command
38
4 Service Port