Owner manual

Syntax Description
li<cr>
Enter values for offset and gain parameters
for user adjustment. Useful for restoring
some earlier state of user adjustment.
Use this command only to
restore values you have
previously written down based
on the output from the l
command, or to restore the
default offset and gain.
Example (shows adjustment has been applied to humidity measurement offset, overwrites it
with 0):
li
Cp offset : -1.50922060E-01 ? 0
Cp gain : 1.00000000E+00 ?
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
Table 35 Li Command
Other Commands
Syntax Description
filt [f.fff]<cr>
Set the speed at which the latest
measurement result is integrated into the
humidity and temperature readings. The
command affects both analog output and
serial line output.
f.fff = Measurement filtering setting, range
0.001 ... 1.0.
n 1.0 = No filtering, latest measurement is
output without averaging
n 0.5 = Average of last two measurements
n 0.1 = Average of approximately 16
measurements
filt<cr>
View the current setting. Also prompts you to
enter a new value.
Example (view the current value and set filtering to value 0.5):
filt
Filter : 1.000 ? 0.5
Table 36 Filt Command
39
4 Service Port