Owner manual

5 MAINTENANCE
Cleaning
The body of the transmitter can be cleaned by wiping with a moistened lint-free
cloth. Do not use cleaning agents or solvents, or blow pressurized air into the
transmitter housing or on the filter.
Do not attempt to clean contaminated HUMICAP
®
sensors or
filters. Instead, replace them with new parts. Filters and sensors
can be purchased from Vaisala. For order codes, see Spare
Parts and Accessories on page59.
Calibration and Adjustment
Calibration means comparing the instrument to a known
reference, either against a second instrument or a known
reference environment. Correcting the reading of the instrument
so that is measures accurately is referred to as adjustment.
Performing an accurate calibration and adjustment takes some
time and preparation. Instead of doing it yourself, you can also
have a Vaisala service center calibrate and adjust your
transmitter. For contact information, see section Product returns
on page55.
HMDW110 series transmitters are fully calibrated as shipped from factory. You
can use the service port to calibrate and adjust the humidity and temperature
measurement of the transmitter as needed. If you think the transmitter is not
measuring correctly, check the following before starting any calibration and
adjustment procedure:
41
5 Maintenance