n The second point (high point) affects measurement gain. For humidity
adjustment, the high point must be > 50 %RH.
Additionally, two-point humidity calibration requires that the difference between
the two points is 30 %RH.
Temperature adjustment points must be in range -40...+60°C (-40 ... +140 °F)
with more than 30 °C (86 °F) difference.
Calibration and Adjustment Using a Hand-Held
Meter and a Reference Probe
n A fully charged Vaisala MI70 indicator (the measurement
display that is included in the HM70 package)
n A calibrated reference probe that is compatible with the MI70
indicator (for example, HMP75) and its MI70 connection
cable, if not integrated with the probe
n Connection cable for HM70 hand-held meter (219980SP)
n Medium size cross-head screwdriver (Pozidriv).
1. Insert the reference probe in the same environment as the probe of the
HMDW110 series transmitter. You can also perform this procedure in the
ambient environment, as long as it is reasonably stable.
2. Connect the reference probe to port Iof the MI70 indicator (the
measurement display that is included in the HM70 package).
3. Prepare the HMDW110 series transmitter for a service port connection as
instructed in section Connecting to the Service Port on page22.
4. Plug in the connection cable 219980SPto the service port of the HMDW110
series transmitter, and the other end to port IIof the MI70 indicator.
5. Turn on the MI70 indicator.
6. Check and adjust the environment settings of the reference probe if
prompted by the MI70.
7. Select Functions > Adjustments (II) from the menu of the MI70 and select
Start. Make sure the symbol for port IIis shown after the word
"Adjustments", as your reference probe may also be adjustable using the
MI70.
8. The adjustment mode starts. Select OK.