M210225en-A Appendix 3: RS 485/422 serial port module
113
NOTE 1
The SMODE command must be given last.
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.
Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
Disconnect the terminal.
Check that the jumpers in connector X4 are in the correct places:
X4
X4
X1 X2
Remount the RS 485/422 serial module.
Close the cover.
Repeat this setting procedure with each transmitter.
When all transmitters on the network have been configured, switch
them off.