Instruction Manual

HMP230 SERIES
M210225en-B Appendix 9: Re-gaining
149
Use arrow keys to change the status of the automatic re-gaining and press
ENT.
Acknowledge the current setting by pressing ENT. To change the interval
press CL. Using arrow keys select the new time interval digit by digit and
acknowledge each digit using ENT key. Using CL key you can correct the
entry by deleting the digits one by one. When you have entered the whole time
interval, press ENT for the second time.
With a similar manner acknowledge or change also the other re-gaining
parameters.
When all the parameters have been changed or acknowledged, the menu
Regen. settings/Regeneration is displayed again.
4.2. Starting re-gaining manually
To start re-gaining manually select Regeneration and press ENT. The
following is displayed:
When the heating is finished or it has been aborted, the settling period begins
and the transmitter returns to the display mode outputting the measured or
calculated values. However, during the settling time or until the temperature
difference is small enough, the display is locked to the values measured before
re-gaining was started. A new re-gaining cannot be started during the settling
period.